eutech-instruments ecoscan do 6 do meter User Manual
Instruction Manual
DO 6
46
7.6.2 User
Reset
This program clears all data including
calibration and other customized setup
functions and reset it to factory default.
calibration and other customized setup
functions and reset it to factory default.
You can skip Calibration Reset and
proceed straight to User Reset
From “rSt.5” menu,
1) Press HOLD/ENTER key three times
until meter displays the User Reset
menu “b.USR”
menu “b.USR”
2) Press HOLD/ENTER key to enter the
menu.
3) Use
the
▲ and ▼ keys to activate or
deactivate the User Reset selection.
a. YES – activates User Reset
b. NO – deactivates User Reset
4) Press HOLD/ENTER key to confirm
the selection made.
5) If selection is ‘YES’, meter would
immediately return to measurement
mode. Otherwise it will return to the
main group menu. Press CAL to return to the measurement mode.
mode. Otherwise it will return to the
main group menu. Press CAL to return to the measurement mode.
Figure 27: User Reset Selection
HOLD
ENTER
User Reset