TROUBLESHOOTING HINTS 
 
Symptom                 Possible Causes          
 
Next Step 
Out of Range            defective meter           
 
check meter with shorting 
Reading                                      
 
 
 
strap (see meter  
                                                
 
 
 
instruction manual) 
 
 
               
defective electrode                 
check electrode operation 
                                    electrodes not          
 
 
unplug electrodes and reset  
                          
plugged in properly    
 
                         
reference electrode       
 
be sure reference electrode 
                         
not filled              
 
 
is filled 
 
                         
air bubble on membrane          
 
remove bubble by 
                         
            
 
 
 
re-dipping electrode 
 
                         
electrodes not in solution 
 
put electrodes in solution 
                       
 
Noisy or Unstable       defective meter           
 
check meter with shorting 
Readings (readings                              
 
 
strap 
continuously or 
rapidly changing)       air bubble on membrane          
remove bubble by  
               
 
 
 
 
 
 
re-dipping electrode 
 
                         
electrode exposed        
 
soak electrode in 
                         
to interferences          
 
cyanide standard 
 
                         
ISA not used           
 
 
use recommended ISA 
 
                         
meter or stirrer not      
 
ground meter or stirrer 
                       
grounded    
 
                         
defective electrode       
 
replace electrode 
 
Drift (reading         
samples and standards   
 
allow solutions to come to 
slowly changing         at different            
 
 
room temperature before 
in one direction)       temperatures              
 
measurement 
 
                         
electrode exposed        
 
check section entitled 
                         
to complexing agents                  
Complexation 
                                              
                         
incorrect reference       
 
use recommended 
                         
filling solution        
 
 
filling solution