eutech-instruments alpha ph differential probe User Manual

Page of 24
Instruction 
manuals 
 
         ECFP21A1A1 
 
 
 
 
7
3.2.1.  Pipe Tee Mounting 
 
Eutech Instruments’ distinct body type for its 1½ inch differential pH sensor ensure optimal 
positioning of the active pH sensor electrode in the process flow.  Normally available pipe tees 
vary widely in construction dimensions.  CPVC pipe tees are most generally available with 
schedule 80 walls.  Stainless steel tees have thinner walls that change the final position of the 
sensor. 
 
Common mounting arrangements are shown for the short nose pH sensors. 
 
 
Recommended Tee Mounting Configurations  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 3.3: Short Nose Sensor Mounted 
in Steel Tee or Schedule 40 PVC Tee 
 
Optimal electrode placement with 
minimum flow restriction 
Dimensions: IN (mm)