eutech-instruments alpha ph differential probe User Manual

Page of 24
Instruction 
manuals 
 
         ECFP21A1A1 
 
 
 
 
15
pH Offset Test 
 
1. Disconnect the RED and GREEN sensor wires from the analyzer terminal block while leaving the other 
wires connected.  Connect the (+) lead of a millivolt meter to the RED wire and the (-) lead of the millivolt 
meter to the GREEN wire.   
 
2. Put the sensor in 7 pH buffer.  Stir the sensor for 10 to 15 seconds and then allow the sensor to stabilize in 
solution for about 2 minutes.  The meter should read 0.0mV +/- 50 mV.  If the reading is higher or lower 
than specified, replace the salt bridge and filling solution and try again.  If on the second attempt the offset 
does not meet specification, replace the sensor. 
 
pH Span Test 
 
If the RED and GREEN wires have already been disconnected to run the offset test, skip step 1. 
 
1. Disconnect the RED and GREEN sensor wires from the analyzer terminal block while leaving the other 
wires connected.  Connect the (+) lead of a millivolt meter to the RED wire and the (-) lead of the millivolt 
meter to the GREEN wire.   
 
2. Rinse the sensor in clean water.  Put the sensor in 4 pH buffer.  Stir the sensor for 10 to 15 seconds and 
then allow the sensor to stabilize in solution for about 2 minutes.  The meter should read a change of at 
least 150mV from the offset reading.  For example, if the offset is +20mV the reading in 4 pH buffer should 
be at least 170mV. 
 
After testing offset and span, the RED and GREEN wires should be reconnected to the analyzer terminal 
block. 
 
Sensor Power Test 
 
If the sensor does not respond with any reasonable millivolt signal in the offset and span tests, it may be 
worthwhile to verify that power is being supplied to the sensor from the analyzer.  To do this, connect a 
voltmeter from the BLACK lead to the WHITE lead (It is not necessary to disconnect the wires from the 
terminal block).  The meter should read –5 VDC +/- 0.3 VDC.  If the voltage is low, disconnect the white wire 
from the terminal block and measure the voltage on the terminal block.  If the voltage comes back into 
specification there is a defect in the sensor that is loading down the analyzer.  If the voltage remains low then 
the analyzer may be defective.    
 
If the sensor passes the temperature, offset, span and power tests but the analyzer is not displaying the 
correct values, the analyzer may be defective.