eutech-instruments bromide epoxy User Manual

Page of 19
Instruction Manual 
 
Bromide Electrode 
 
17
TROUBLESHOOTING HINTS 
 
Symptom                 Possible Causes          
 
Next Step 
Out of Range      
defective meter           
 
check meter with shorting 
Reading                                      
 
 
 
strap (see meter  
                                               
 
 
 
instruction manual) 
 
                    
 
defective electrode              
 
check electrode operation 
                       
                         
electrodes not    
       
 
unplug electrodes and reset 
                           
plugged in properly     
 
                         
reference electrode       
 
be sure reference electrode 
                         
not filled              
 
 
is filled 
 
                         
air bubble on membrane 
     
remove bubble by replacing 
                    
 
 
 
 
 
 
electrode 
 
                        
electrodes not in solution 
 
put electrodes in solution 
                       
 
Noisy or Unstable       defective meter           
 
check meter with 
Readings (readings                              
 
 
shorting strap 
continuously or 
rapidly changing)       air bubble on membrane          
 
remove bubble by  
                    
 
 
 
 
 
 
re-dipping electrode 
 
                         
electrode exposed        
 
soak electrode in 
                         
to interferences          
 
bromide standard 
 
 
               
defective electrode 
 
 
replace electrode 
 
                         
ISA not used              
 
use recommended ISA 
 
                         
meter or stirrer        
 
 
ground meter or stirrer 
                        
not grounded  
   
 
Drift (reading         
samples and standards   
 
allow solutions to come 
slowly changing         at different temperatures            
to room temperature 
one direction)     
 
 
 
 
 
before measurement 
 
                         
electrode exposed        
 
check section entitled 
                         
to complexing agents                  
Complexation 
                                              
                         
incorrect reference       
 
use recommended 
                         
filling solution        
 
 
filling solution