dhi ppcmpc vacuum kit instructions pn 400922 (110v) pn 401160 (220v) User Manual

Page of 1
 
Calibration Solutions 
for Pressure 
and Flow™ 
PPC
/MPC
 Vacuum Kit 
Instructions 
P/N 400922 (110V) 
P/N 401160 (220V)
 
THE 110V VACUUM KIT INCLUDES 
•  1 
102381  [102034] 
Fast Vac Vacuum Pump 
•  1 
101322  [102382] 
3-Way Solenoid Valve 
•  1 
102123 
Adaptor AN4F Swivel x 1/8 in. NPTM 
•  2 
102033 
Adaptor 1/4 in. Swage x 1/8 in. NPTM 
•  1 
100109 
Adaptor 1/4 in. Swage x 1/4 in. NPTM 
•  1 
101450-Z 
1/4 in. O.D. Tubing (200cm) 
• 
The [ ] above denotes the part number required for a 220V Vacuum Kit 
100109  or
 
The PPC/MPC Vacuum Kit should be installed as illustrated above.  Fill the vacuum pump with oil before using.  
Read Fast Vac instruction sheet for additional information. 
A three-way solenoid valve has been installed to protect the vacuum pump when turned off.  When decreasing 
pressure in a PPC the test gas is exhausted through the reference port.  This ensures that the gas will not 
exhaust into the pump and possibly cause damage.  The valve is electrically connected to the vacuum pump 
power switch and will be actuated when the pump is turned on. 
When the vacuum pump is turned off, the reference or vacuum port of the PPC (or MPC) will be connected to 
atmosphere bypassing the vacuum pump. 
When the vacuum pump is turned on, the reference port of the PPC (or the vacuum port of the MPC) will be 
connected to the vacuum pump enabling pressure below atmosphere to be generated. 
PPC, RPM and MPC are trademarks, registered and otherwise, of DH Instruments, Inc.
 
Document 560004b 
040317 
DH Instruments, Inc. 
4765 East Beautiful Lane 
Phoenix  AZ  85044-5318 
USA 
 
 
Tel 602.431.9100
Fax 602.431.9559
dhi@dhinstruments.com
www.dhinstruments.com