Tektronix DMM4020 Digital-Multimeter, DMM, VFD-Multiple segment, 20000 counts CAT II 600 V DMM4020-A1 Data Sheet
Product codes
DMM4020-A1
Operating the Meter from the Front Panel
Dual Display
3
3-7
Table 3-2. Display Annunciators and Indicators
No. Name
Description
A
r
Meter is in remote mode (remotely controlled)
B
L
The front panel is locked out
C
Ext Trig
Meter is in external trigger mode
D
S
Q
is pressed and secondary function will be selected
E
REL
X
Relative readings function modifier is selected
F
–
Measurement value in secondary display is negative
G
H
Touch Hold function modifier is selected
H
M
Minimum maximum function modifier is selected
I
MAX and MIN
Reading is maximum or minimum
J
AC DC
Form of voltage displayed in secondary display
K
2x4 Wire
Resistance measurement method selected (2-wire or 4-
wire)
wire)
L
mV µmA kHz Mke
Unit of measurement displayed in secondary display
M
R
G
Continuity test or diode test is selected
N
mV µmA Mke kHz
Unit of measurement displayed in primary display
O
AC DC
Form of voltage displayed in primary display
P
Slow, Medium, Fast
Measurement rate selected (slow, medium, fast)
Q
–
Measurement value in primary display is negative
R
Setup 8 annunciator
Which configuration is currently selected
S
Print
Meter is in RS-232 print-only mode
T
* (asterisk)
Flashes for each Meter sample cycle
U
Auto Range
Meter is in auto range mode
V
Manual Range
Meter is in manual range mode
W
CAL Error
Calibration attempt failed
X
Y
High voltage is detected
Displays when voltage is >30 V dc or ac rms