  SYSTEM/DISPLAY CONFIGURATION 
81 
Compatibility Settings 
Changing the Compatibility Setting 
The GDM-8341/8342 can be set to a special compatibility mode that 
will allow the unit to emulate the SCPI command syntax of the 
GDM-8246 when in remote control mode. For example, this feature 
can allow programs that were originally written for the GDM-8246 
to run on the GDM-8342/8341 with little modification.  
 
Range 
 LANG 
NORM, COMP 
Procedure 
1.  Press the MENU key. 
2.  Go to SYSTEM on level 1. 
3.  Go to LANG on level 2. 
4.  Set the LANG setting to either NORM (normal 
mode) or COMP (compatibility mode). 
5.  Press the Enter key to confirm. 
6.  Press the EXIT key to exit from the LANG 
menu. 
Display 
LANG setting