Greisinger GOX 100 T Oxygen Meter with Sensor 600222 Data Sheet
Product codes
600222
30
WK
Handheld instrument
Display / Controller
Logger / EASYBus
Transmitter
Temperature probe
Alarm / Protection
Air oxygen measuring device
Double display for oxygen and temperature
Measured units: O
2
-concentration and
O
2
-partial pressure
Alarm detector with integrated horn
Automatic temperature compensation
Min./Max. value memory, Hold function
Serial interface, device can be connected
to bus system (up to 5 devices can be con-
nected to one PC interface)
Battery and d.c. operation
Wide range of application
Most simple calibration in atmospheric air
Calibration: 1-point calibration: extremely
simple quick calibration in atmospheric air. (press
button to compensate unit to 20.9%).
2-point calibration: fi rst point at atmospheric air
(20.9%), second point freely selectable
Application: Wide range of application for your
(20.9%), second point freely selectable
Application: Wide range of application for your
home, job and hobby! For example:
- Bio chemistry: Oxygen monitoring in breeding
chambers for cell cultures. Monitoring of fermen-
ting process of fruits in fermentation plants etc.
- Medicine: Monitoring of oxygen concentration
in respirators; checking of breathing, monitoring
of oxygen concentration in incubators, oxygen
tents etc.
- Food technology: Monitoring of residual oxygen
in packages (e.g. coffee, tea, etc.). Monitoring of
oxygen content during production processes.
- Safety technology, safety at work:
Oxygen monitoring in mines/pits, underground
parking lots, wine cellars, cooling chambers,
greenhouses or stores. Oxygen monitoring or
alarm in case of danger of suffocation when
working in tanks, wells etc.
- Air conditioning and ventilation technology:
Oxygen measurements, air quality monitoring,
measuring of oxygen concentration in enclosed
air conditioning systems, etc.
- Sport: Checking of oxygen content in com-
pressed air breathing apparatuses (diving, etc.),
oxygen monitoring for gliding.
The device can only be used to check during
The device can only be used to check during
these applications. -> no substitute for ap-
proved monitoring device!
Specifi cation:
Measuring ranges:
Oxygen concentration: 0,0 ... 100,0 % O
2
(gaseous)
Partial oxygen pressure: 0 ... 1100 hPa O
2
Temperature: -5,0 ... 50,0 °C
Accuracy:
(device) (at nominal temperature = 25°C)
Oxygen concentration: ±0.1% ±1digit
Partial oxygen pressure: ±1 hPa ±1digit
Temperature: ±0.1°C ±1digit
Oxygen electrode: for suitable sensores
p.r.t. page 31
Sensor connection: 6-pin screened Mini-DIN-
socket.
Display: two 4 digit LCDs
(12.4mm or 7mm high)
,
as well as additional arrows.
Pushbuttons: 6 membrane keys for ON/OFF-
switch, selection of meas. range, min- and max-
value memory, hold-function, calibration etc.
Working temperature: 0 to +50°C
Relative humidity: 0 to +95%RH
(non-condensing)
Storage temperature: -20 to +70°C
Interface: serial interface,
direct connection to RS232 or USB interface of
a PC via electrically isolated interface converter
GRS3100 or GRS3105 resp. USB3100 (p.r.t.
accessories).
Power supply: 9V-battery, type IEC 6F22
(in-
cluded),
as well as additional d.c. connector for
external 10.5-12V direct voltage supply. (suitable
power supply: GNG10/3000)
Power-Off-function: 1...120min (can also be
Power-Off-function: 1...120min (can also be
deaktivated).
Power consumption: approx. 1.5 mA
Power consumption: approx. 1.5 mA
Low battery warning:
and ' bAt '
Dimensions: 142 x 71 x 26 mm (H x W x D)
Impact-resistant ABS plastic housing, membrane
keyboard, transparent panel. Front side IP65,
integrated pop-up clip.
Weight: approx. 160 g (cpl. with battery)
Functions:
Min-/Max-value memory: max. and min. values
will be memorized.
Hold function: by pressing a button the current
meas. value will be memorized.
Alarm: integrated limit detector for min. or max.
alarm.
Temperature compensation: automatic via tem-
perature sensor, integrated in probe housing.
Air pressure compensation: The O
2
concen-
tration will be compensated according to the abs.
atmospheric pressure set (500...2000hPa).
GMH
3691
Sensor not included - please order separately!
Accessories:
Suitable sensores
p.r.t. page 31
GKK 3000
case
(275 x 229 x 83 mm)
with punched lining suitable for GMH3xxx
GRS 3100
interface converter,
electrical isolated, for RS232
GRS 3105
interface converter
with 5 connection points, electr. isolated, for the
connection of 5 GMH3xxx to one PC (RS232).
ST-R1
device protection bag
with cut-out for probe connection
for add. accessories p.r.t. pages 39 - 41
Compact
air oxygen meas. device
Spare peaces, accessories:
GOEL 369
spare sensor for GOX 100
GOEL 370
spare sensor for GOX 100T
ESA 369
spare tube-adapter
ZOT 369
spare T-piece
GKK 252
case
(235 x 185 x 48 mm)
with foam lining
for add. accessories p.r.t. page 40/41
GOX
100
for universal applications
1-Button Calibration
Automatic Power-Off
Min-/max- value memory
Incl. sensor GOEL 369
GOX
100T
for diving applications
1
-Button Calibration
MOD-Display
(Maximum Operating Depth)
HOLD function
Incl. sensor GOEL 370
•
•
•
•
•
•
•
•
Specifi cation:
Meas. range: 0,0 ... 100,0 % O
2
Accuracy: ± 0,1 % O
2
± 1 digit
Sensor Connection: jack-connector cable
Sensor: Oxygen-partial pressure probe,
mounted in external sensor housing
Warranty:
12 months
Working pressure: 0,5 to 2,0 bar absolute
Over-/under-pressure: max. 0,25 bar
Working temperature: 0 to 50°C (sensor GOX 100)
0 to 45°C (sensor GOX 100T)
-20
to
50°C
(device)
Relative humidity: 0 to +95%RH
Power supply: 9V battery type IEC 6F22
Power consumption: approx. 120µA
(over 2500 h)
Display: 3½-digit, 13mm high LCD-display
Housing: ABS-enclosure, front side IP65
Dimensions: approx. 106 x 67 x 30 mm
Weight: approx. 185g
Features:
BAT,
Auto-Power-Off
Scope of supply:
Device incl. sensor, T-piece, fl ow diverter
Options:
- LACK
encapsulated PC board
(for applications where condensation is possible)