Greisinger Oxy 3690 Oxygen Meter 602951 Data Sheet

Product codes
602951
Page of 6
T02.0.4X.6C-09 
 
page 3 of 6  
Assignment of elbow-type plug: 
 
2-wire connection (4-20mA) 
3-wire connection (voltage) 
4-wire connection (voltage) 
 
 
1  =  supply voltage +Vs 
=  signal + 
=  signal + 
 
2  =  GND / signal 
=  supply voltage +Vs 
=  signal - 
 
   
 
(4)  =  supply voltage -Vs 
=  supply voltage +Vs 
 
   
 
    signal - 
 
(4)  =  supply voltage -Vs 
General installation instructions: 
To mount the connection cable (2-, 3-, or 4-wire depending on type of device) the elbow-type plug screw has to be loosened 
and the coupling insert has to be removed by means of a screw driver at the position indicated (arrow).   
Pull out connection cable through PG glanding and connect to the loose coupling insert as described in the wiring diagram. 
Replace loose coupling insert onto the pins at the transducer housing and turn cover cap with PG glanding in the direction 
desired till it snaps on (4 different starting positions at 90° intervals). Re-tighten the screw at the angle plug. 
 
Display functions: 
During normal operation the oxygen content of the air is displayed in the unit [%]. By pressing the Key 2 (down) the 
temperature at the sensor can be shown, pressing key 3 (up) shows the electrode rating in [%]. The electrode rating is 
calculated and stored during the calibration (see below). The displays of the rating and the temperature of the sensor are 
marked by small arrows at the upper end of the display. After 5 seconds the transducer display switches automatically 
back to the oxygen value. 
 
 
 
display dissolved oxygen 
display sensor temperature 
display electrode rating 
 
Calibration of the sensor: 
Expose sensor to the ambient air, wait until the temperature has adopted to the ambient conditions. 
Press key "SET" for 2 sec., The display shows "CAL". 
After approx. 10 sec‟s the transducer is calibrated or a referring error message will be shown: 
Display  Meaning 
Possible fault causes 
Remedy 
CFE.1 
temperature out of allowed 
range  
temperature has to be within 5 to 
40°C  
calibrate again at correct tempera-
ture 
sensor error 
temperature sensing defective 
check cable & connection,  
replace sensor if necessary 
CFE.3 
wrong signal: to low 
sensor element used up 
replace sensor element 
CFE.4 
wrong signal: to high 
calibration surrounding not valid  
check Calibration surrounding 
CFE.6 
instable signal 
calibration surrounding not valid  
check Calibration surrounding 
During calibration the rating of the electrode is evaluated: 100% means perfect state, 40 % means e.g. sensor element 
has a weak output signal and needs to be replaced soon (p.r.t. sensor manual). The electrode rating can be displayed 
with key 3 (down).