Instruction Manual
EL 9000 HP Series
Date: 10-31-2011
Handling the device 
Rotary knob 
Setting (6)
This rotary knob has no end stop. With every 
position 
Setting (6) changes the set value 
resp.  the  parameter  (in  the  setup  menu) 
which was chosen before by 
Selection (5).
It  applies:  the  faster  the  knob  is  rotated, 
the faster the set value is increased or decreased (the 
step width changes). Slow rotation hence changes the 
set value in small steps, while fast rotation changes it in 
big steps.
In case the upper or lower limit of the set value is reached 
while rotating, the set value will remain at its limit until 
it  changed  again  by  rotating  the  knob  in  the  opposite 
direction.
6.3  Switching power on
The device is powered by switching the mains switch on. 
Then the device type and firmware version are shown 
on the display
After the internal system has booted and performed some 
tests, the former set values are restored, if setting 
Keep
set values
 is set to 
yes
,or default values are set.
The load‘s DC input will always be offline after the first 
start.
6.4  Switching the DC input on/off
By pressing the pushbutton 
Input on/off (4) the DC input 
of the load input is switched on and the device starts to 
work as a load. 
Pressing the button again switches the input off.
The input can also be switched off with the pin named 
„REM-SB“ on the analogue interface (pin = low), if it was 
on before. Switching the input on if it was off before is 
not possible.
While the load is working, all four actual values of current, 
voltage, power and resistance are indicated on the left 
half of the display.
When the load input is off, only the actual value of volta-
ge is indicated, because there are no reasonable other 
actual values available at this moment. In battery test 
mode, the time counter is also halted resp. continued if 
the pushbutton is pressed.
The activation of the load operation can be inhibited by 
certain reasons. For example, if an overvoltage or another 
error is persistent or the load is remotely controlled by 
the analogue or digital interface.
The standby state (input = off) is displayed like this:
Figure 8
6.5  Adjusting the set values
Note: Set values, which must not be changed in the cur-
rently selected regulation mode, are not shown.
Set values can be selected and adjusted as long as the 
display does not show the status texts 
External mode
 
or 
Remote mode
.
a) Selecting the set value to adjust
In the operation modes 
ABA/B and Battery the set 
value to change is selected by rotating 
Selection (5)
In the setup menu (position 
Setup) it is used to select a 
different setup parameter. Also see „7.1. The setup menu“.
b) Adjusting the set value
The previously with 
Selection (5) selected set value is 
changed with 
Setting (6). It applies: the faster the knob is 
rotated, the faster the set value is increased or decreased 
(the step width changes). Slow rotation hence changes 
the set value in small steps, while fast rotation changes 
it in big steps.
In case the upper or lower limit of the set value is reached 
while rotating, the set value will remain at its limit until 
it  changed  again  by  rotating  the  knob  in  the  opposite 
direction.
Important! Note, that in Level A/B mode any set value
of Level A must always be greater than or equal to
Level B! The result of this is, that Level B can upwards
only be adjusted to the value of Level A and Level
A can downwards only be adjusted to the value of
Level B.
6.6  Preselecting the regulation mode
The selector 
Mode (2) preselects the regulation mode. 
Four  modes  are  available: 
CC,  CV,  CP  and  CR. The 
set values will later determine the true behaviour of the 
device.
CC
means constant current. Here are the set values 
of current and power adjustable. In this mode the input 
current  is  regulated  and  limited  to  the  adjusted  value 
(current limitation), as far as the feeding source is able to 
deliver the current. The set value of power is additionally 
limitting the consumed power and additionally effects the 
maximum current. The power limitation is dominant over 
the current limitation.
A changeover to regulation mode 
CC can reset the set 
value of power to its nominal value and the set value of 
current to 0, if the parameter 
Keep set values
 has been 
set to 
no
 in the setup. If set to 
yes
, the least adjusted set 
CV
 means  constant  voltage.  Here  the  set  values  of 
voltage, current and power are adjustable. In this mode 
the input voltage is limited to the adjusted set value by 
loading the feeding source so much that its output voltage 
collapses. If the feeding source is able to deliver more 
current  than  the  device  is  able  to  withdraw,  then  the 
voltage limit can not be achieved.