Gw Instek GSP-830Spectrum analyzer, spectrum analyzerBandwidth (RBW) 3/30/300 kHz/4 MHz GSP-830 Data Sheet

Product codes
GSP-830
Page of 181
 
 
GSP-830 User Manual 
128
 
 
3.  Press F2 (Self  Test). 
Self Test...
F   2
 
 
4.  The GPIB result appears at F1. 
If  the result is Fail (underlined), 
contact service personnel. 
GPIB
Pass   Fail
F   1
 
 
GPIB constraints 
 
Keep these rules when using the GPIB interface. 
• 
Altogether less than 15 devices & 20m cable length, 
2m between each device on the bus 
• 
Unique address assigned to each device 
• 
At least 2/3 of  the GPIB devices turned on 
• 
No loop or parallel structure allowed 
View System Information 
View system error 
 
Panel operation 
View the bottom of  the screen, the error message area. 
If  there is a system error, the message appears in red. 
Center : 1.5GHz
(EXT Unlock)
EXT
Unlock
 
 
Amp
Uncal
 
• 
Inappropriate RBW or VBW is selected. 
• 
Frequency is less than 15MHz and 
amplitude is less than −30dBm. 
 
EXT
Unlock
 
External reference input is not working 
properly. 
 
LO1
Unlock
 
Local oscillator 1 is not working properly. 
 
LO3
Unlock
 
Local oscillator 3 is not working properly. 
 
Med
Unlock
 
±
1ppm stability signal is not working 
properly. Appears only when the optional 
±
1ppm Stability Module is installed. 
 
Ref
Unlock
 
Internal reference signal is not working 
properly.