Kern GAB 12K0.1NParcel scales Weight range bis 12 kg GAB 12K0.1N User Manual
Product codes
GAB 12K0.1N
18
GAB-N-BA-e-1315
9.2 Non-verifiable models:
Switch on balance Whilst balance is carrying out self-test
(counts down from 99… to 00…)
TARE
and
%
press at the
same time.
“UnLoad“ will be displayed, follo
wed by flashing numeric
value.
(Example)
Change by pressing the navigation keys (See manual chap.
8); select desired setting; currently enabled digit will be
flashing.
(Example)
Confirm value by
0
, “Load“
will be displayed.
Place adjustment weight, stability display appears
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during self test,
balance will return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting weight
the display will show an error message, repeat adjustment
process.
GROSS
ZERO
ZERO