Kern GAB 6K0.05NParcel scales Weight range bis 6 kg GAB 6K0.05N User Manual

Product codes
GAB 6K0.05N
Page of 41
GAB-N-BA-e-1315 
 
17 
9.1  Verified models 
 
  Switch on device via ON-OFF and press the adjustment 
switch at the same time. 
  After that, during the balance’s selftest, press 
TARE
 and 
%
 
at the same time. „UnLoAd“ appears. 
 
 
 
  Acknowledge by 
. Ensure that there are no objects on 
the weighing plate. 
 
 
 
  The currently set adjustment weight will be displayed. 
  To change by using the navigation buttons (see operating 
instructions chap. 8) select the desired setting, the active 
digit is flashing.   
 
 
 
 
  Acknowledge by 
.  „LoAd“ is displayed. 
 
  Carefully place adjusting weight in the centre of the  
weighing plate. Wait for stability display, then press 
„PASS“ appears
 
 
 
 
  After the adjustment the balance will carry out a self-test. 
Remove adjusting weight during selftest, the appliance will 
return into weighing mode automatically.  
An adjusting error or incorrect adjusting weight will be 
indicated by the error message; repeat adjustment 
procedure. 
GROSS
ZERO