Linearity shows the greatest deviation of a weight display on the scale to the value of 
the respective test weight according to plus and minus over the entire weighing 
range. 
If linearity deviation is discovered during a testing instrument control, you can 
improve this by means of linearization. 
 
 
  Carrying out linearization is restricted to specialist staff possessing 
well acquainted with the workings of weighing scales. 
  The test weights to be used must be adapted to the weighing scale’s 
specifications; see chapter 2.4 “Testing instruments control”. 
  Observe stable environmental conditions. Stabilisation requires a 
certain warm-up time. 
  After successful linearization you will have to carry out calibration; 
see chapter 2.4 “Testing instruments control” 
 
Tab. 1: Adjustment points
 
 
Adjustment 
weight 
GAB 6K0.05N  
GAB 12K0.1N
 
GAB 30K0.2N 
1.   
0 kg 
0 kg 
0 kg 
2.   
2 kg 
4 kg 
10 kg 
3.   
4 kg 
8 kg 
20 kg 
4.   
6 kg 
12 kg 
30 kg