GAB-N-BA-e-1315 
 
12  Data output 
 
12.1  RS232 interface 
 
The RS 232 interface allows a bi-directional data exchange from the balance to 
external devices. This data exchange is asynchronous using ASCII - Code. 
 
12.1.1  Technical data 
 
  ASCII code 
  8 data bits 
  Baud rate selectable at 600, 1200, 2400, 4800, 9600 bps  
  Miniature plug-in necessary (9 pole D-Sub) 
  No parity 
  For  operation  with  interface  faultless  operation  is  only  ensured  with  the  correct 
KERN – interface cable (max. 2m) 
 
12.1.2  Pin allocation of balance output bushing: 
 
 
 
 
      1
 
 
                            
9             6    
 
 
Pin 2:  Output 
Pin 3:  Input, not used 
Pin 5:  Signal ground