Kern GAB 6K0.05NParcel scales Weight range bis 6 kg GAB 6K0.05N User Manual

Product codes
GAB 6K0.05N
Page of 41
40 
 
GAB-N-BA-e-1315 
 
14  Instant help 
 
In case of an error in the program process, briefly turn off the balance and disconnect 
from power supply. The weighing process must then be restarted from the beginning. 
Help: 
Fault 
Possible cause 
 
 
The displayed weight does 
not glow. 
  The balance is not switched on. 
 
  Mains power supply interrupted (mains cable 
defective). 
 
  Power supply interrupted. 
 
  Rechargeable battery inserted incorrectly or empty 
 
 
The displayed weight is 
permanently changing 
  Draught/air movement 
 
  Table/floor vibrations 
 
  Weighing plate has contact with other objects. 
  Electromagnetic fields / static charging (choose 
different location/switch off interfering device if 
possible) 
 
 
The weighing result is 
obviously incorrect 
  The display of the balance is not at zero 
 
  Adjustment is no longer correct. 
 
  Great fluctuations in temperature. 
  Warm-up time was ignored. 
  Electromagnetic fields / static charging (choose 
different location/switch off interfering device if 
possible)  
 
 
 
Should other error messages occur, switch balance off and then on again. If the error 
message remains inform manufacturer.