Kern FFN 15K2IPNParcel scales Weight range bis 15 kg FFN 15K2IPN User Manual

Product codes
FFN 15K2IPN
Page of 31
FFN-N-BA-e-1222 
 
20 
 
7.1.2  Adjusting verifiable models 
 
 
The adjustment is locked for verified balances.  
with adjustment switch 
To disable the access lock, destroy the seal and actuate the adjustment 
switch. Position of the adjustment switch see chapter 7.2.1 
Attention
After destruction of the seal the balance must be re-verified by an 
authorised agency and a new verification wire/seal mark fitted before it can 
be reused for applications subject to verification. 
 
 
Display 
Operation 
 
 
 
  Start balance by pressing 
 
  While the balance 
carries out a self test (00…- 99…), press 
 
and keep pressed until „F1 CAL“ appears on the 
display. 
  Actuate the adjustment switch on the lower side of the 
balance 
 
 
 
   Press 
 
in the display appears „UnLod“. 
 
 
 
(Beispiel) 
   Press 
 anew 
The weighing value is displayed. 
 
 
 
 
 
 
 
  Press 
 to enter the value of the adjustment weight 
(see chapter 1)  
  Place the adjustment weight 
   Press 
 
While the balance carries out a self-test, take away the 
adjustment weight 
 
The balance changes over to zero display. 
Now the process of adjustment is completed.