5  Unpacking, installation and commissioning 
5.1  Place of installation, place of use 
 
The balance is constructed in such a way that reliable weighing results can be achie-
ved under normal application conditions.  
By selecting the correct location for your balance, you will be able to work quickly and 
precisely.  
 
Therefore please observe the following at the place of installation: 
 
Place the balance on a firm, level surface; 
Avoid extreme heat as well as temperature fluctuation caused by installing 
next to a radiator or in the direct sunlight;  
Protect the balance against direct draughts due to open windows and doors;  
Avoid jarring during weighing;  
Protect the balance against high humidity, vapours and dust;  
Do not expose the device to extreme dampness for longer periods of time. In-
admissible bedewing (condensation of air moisture on the device) can occur if 
a cold device is taken into a significantly warmer environment.  In this case, 
please acclimatise the device for approx. 2 hours at room temperature after it 
has been disconnected from the mains.  
Avoid static charging of the material to be weighed, weighing container and 
windshield.   
 
Major display deviations (incorrect weighing results) are possible if electromagnetic 
fields occur as well as due to static charging and instable power supply. It is then ne-
cessary to change the location.  
 
 
 
5.2  Unpacking the Balance 
 
Carefully remove the balance from the packaging and out of the plastic covering, 
then place the balance on the assigned position. 
 
5.2.1  Installation 
 
Install the balance in such a fashion that the weighing plate is absolutely horizontal.