Kern Parcel scales Weight range bis 60 kg EOB 60K20 User Manual

Product codes
EOB 60K20
Page of 24
7  Adjustment 
 
As the acceleration value due to gravity is not the same at every location on earth, 
each balance must be coordinated -  in compliance with the underlying physical 
weighing principle - to the existing acceleration due to gravity at its place of location 
(only if the balance has not already been adjusted to the location in the factory). This 
adjustment process must be carried out for the first commissioning, after each 
change of location as well as in case of fluctuating environment temperature. To 
receive accurate measuring values it is also recommended to adjust the balance 
periodically in weighing operation.  
Observe stable environmental conditions. A warming up time (see chapter 1) is 
required for stabilization. 
 
Ö  In weighing mode press the 
UNIT key and keep it pressed until „CAL“ followed by 
the weight value of the required adjustment weight in displayed. „
LOAd“ is 
displayed alternatively. 
 
Ö
 
 
Ù
 
 
 
 
 
 
Example 
Ö  Carefully place the required adjustment weight (see tab. 1 „Technical data“) in 
the centre of the weighing plate 
Press 
TARE key, adjustment is started. 
 
Ö
 
 
 
 
 
Ö  Wait until the stability display appears, after that the adjustment is finished. The 
balance will automatically return to Weighing mode upon successful completion 
of the adjustment and will display the weight of the adjustment weight. 
 
Ö  Take away adjustment weight. 
 
 
 
 
An error message will be displayed in the event of an adjustment error or 
incorrect adjustment weight. Remove the adjustment weight and repeat the 
adjustment process. 
 
Keep the adjustment close to the balance. Daily control of the weighing 
exactness is recommended for quality-relevant applications. 
 
EOB/EOE_L/EOE_XL/EOS-BA-e-1030 
17