6.2.6  Reset function 
Factory setting reset. 
Ö  When the balance is switched off, press the OFF -key and keep it pressed 
Ö  Do not release the OFF -key. Press also the ON/TARE-key and keep it pressed  
Ö Keep the ON/TARE-key pressed, however release the OFF -key 
Ö Press OFF -key anew 
Ö  Keep both keys pressed until in the display “tr” appears 
Ö  Release both keys. The balance is situated in the menu. 
Ö Press the ON/TARE-key until „rSt“ appears  
Ö  Confirm with the OFF key  
Ö Press the ON/TAREkey to switch among the settings: 
 
Display 
Mode 
„rst“ YES 
Factory setting reset 
„rst“ NO 
No factory setting reset 
 
Ö Confirm with the OFF key. Wait until the balance automatically returns to weighing 
mode