8  Troubleshooting  
The balance should be switched off for a short time following an interruption in the 
programme sequence and disconnected from the mains supply. It is then necessary 
to repeat the weighing process from the beginning.  
 
Help: 
Interruption Possible 
cause 
 
 
Weight display is not illuminated. 
•  The balance is not switched on. 
 
•  The batteries are wrongly inserted, the 
batteries are empty 
 
•  No  batteries are attached  
 
•  The mains supply connection has been 
interrupted (mains cable not plugged 
in/faulty). 
 
•  Power supply interrupted. . 
 
 
The weight display changes continu-
ally   
• Draught/air movement 
 
•  Table/floor vibrations  
 
•  The weighing plate is in contact with for-
eign matter. 
•  Electromagnetic fields / static charging 
(choose different location/switch off inter-
fering device if possible) 
 
 
The weighing result is obviously in-
correct 
•  The balance display is not set to zero  
 
•  Adjustment is no longer correct. 
 
•  Great fluctuations in temperature. 
•  Electromagnetic fields / static charging 
(choose different location/switch off inter-
fering device if possible) 
 
Switch the balance off if other error messages should appear and then switch on 
again. Contact the manufacturer if the error message does not disappear.