6.3  Battery operation and change 
In order to save the battery, the balance switches automatically off after 30 seconds 
without weighing. 
 
 
If  the  balance  is  not  used  for  a  longer  time,  take  out  the  batteries  and  store  them 
separately. Leaking battery liquid could damage the balance. 
6.4  Initial Commissioning 
The warm-up period after the start stabilises the measured values.  
The accuracy of the balance depends on the local acceleration of gravity.  
Strictly observe hints in chapter “Adjustment”. 
6.5  Adjustment 
As the acceleration value due to gravity is not the same at every location on earth, 
each balance must be coordinated - in compliance with the underlying physical 
weighing principle - to the existing acceleration due to gravity at its place of location 
(only if the balance has not already been adjusted to the location in the factory). This 
adjustment process must be carried out during the initial start-up, after change in 
location and variation of surrounding temperature. To receive accurate measuring 
values it is also recommended to adjust the balance periodically in weighing 
operation. 
6.6  Adjusting 
With an adjustment weight, the weighing accuracy can be checked and re-adjusted 
at any time. 
Procedure when adjusting: 
Observe stable environmental conditions.  
 
   Switch on scale by 
 
and wait for “0.0“ to appear 
on the screen 
 
 
 
  Press 
 
and hold until “ZERO“ appears flashing 
 
 
 
  Confirm by pressing 
 and the value of the 
adjustment weight will be shown flashing 
 
 
 
 
Position the adjustment weight; after a while “PASS“ 
appears; this concludes the adjustment process. 
Afterwards the scale switches off automatically.