Kern DE 300K50DL Postal Scale 300kg DE 300K50DL User Manual

Product codes
DE 300K50DL
Page of 37
DE-BA-e-1255 
15 
 
5.5  Connection of peripheral devices   
Before  connecting  or  disconnecting  of  additional  devices  (printer,  PC)  to  the  data 
interface, always disconnect the balance from the power supply. 
With  your  balance,  only  use  accessories  and  peripheral  devices  by  KERN,  as  they 
are ideally tuned to your balance. 
 
 
5.6  Initial Commissioning 
In order to obtain exact results with the electronic balances, your balance must have 
reached the operating temperature (see warming up time chap. 1). During this 
warming up time the balance must be connected to the power supply (mains, 
accumulator or battery). 
 
The accuracy of the balance depends on the local acceleration of gravity.  
Strictly observe hints in chapter Adjustment. 
 
 
5.7  Adjustment 
As the acceleration value due to gravity is not the same at every location on earth, 
each balance must be coordinated - in compliance with the underlying physical 
weighing principle - to the existing acceleration due to gravity at its place of location 
(only if the balance has not already been adjusted to the location in the factory). This 
adjustment process must be carried out during the initial start-up, after change in 
location and variation of surrounding temperature. To receive accurate measuring 
values it is also recommended to adjust the balance periodically in weighing 
operation. 
 
5.8  Adjustment 
The adjustment should be made with the recommended adjustment weight (see 
chap. 1 "Technical data"). Adjustment is also possible with the weights of other 
nominal values (see table 1), but not the optimum for measuring technique.