SOEKS IMPULSE Magnetic field-analysis apparatus, Magnetic field tester 9 Data Sheet
Product codes
9
Specification
Table 1
Measurable electromagnetic field frequency from 20 to 2 000
range, Hz
Measurement range of magnetic field from 0,04 to 12*
(magnetic induction) intensity amplitude (from 0,05 to 15*)
along X, Y, Z axes, A/m (uT)
Measurement range of magnetic field from 0,08 to 20*
(magnetic induction) intensity mean square (from 0,10 to 25*)
value, A/m (uT)
Maximum permissible relative error of ±30
measurement of magnetic field intensity, %
Hardware nonlinearity of measurement of 7,0
magnetic field intensity in the measurement
range for test magnetic field of 50 Hz in
frequency, not more than, %
Measurement range of electric field from 10 to 1000*
intensity amplitude along X, Y, Z axes, V/m
Measurement range of electric field from 17 tо 1700*
intensity mean square value, V/m
Maximum permissible relative error of ±30
measurement of electric field intensity, %
Hardware nonlinearity of measurement of 4,0
electric field intensity in the measurement
range for test magnetic field of 50 Hz in
frequency, not more than, %
Comment:
* - not less than specified parameter value
* - not less than specified parameter value
-26-