INDICATORS 
 
    HEATING ....... This red indicator is lit when the heating 
 
    element is actually drawing current from the circuit and 
 
    heating the water. When it reaches set temperature, the 
 
    heating indicator light shuts off. 
 
        FAULT ............ If the water temperature exceeds 58 
°C, 
 
    an alarm will sound and the red fault light will be 
 
    illuminated. This may indicate a unit malfunction. 
 
        ADD WATER .... This red indicator will light if there is 
 
    insufficient water in the bath. The unit will not operate. 
 
    Add water to correct this condition. This indicator will 
 
    light also when the shroud unit is lifted in up position. 
 
 
 
 
7.  TROUBLESHOOTING GUIDE  
 
  PROBLEM and ACTION 
 
  1) The indicator lights and displays do not come on when 
the POWER key is pressed. 
   
• Check the POWER SWITCH on the left-hand side. 
   
• Check the outlet for 110/120 VAC 60 Hz. 
   
• Check the fuse in the unit. Replace the fuse if necessary. 
   
• Ensure that the POWER key is pressed for at least one 
   second. 
   
• Check that there is sufficient water in the bath and that 
      the ADD WATER light is off. 
   
• Check that the FAULT light is off. 
 
 
 
15