Genesis Advanced Technologies 2002 User Manual

Page of 22
 
 
 
1.  UNPACKING 
 
1.1   
The following is supplied with your GENESIS 2002/2006
 
  
unit. 
                    Control Unit 
                    Water Bath 
                    Thawing Bag Dispenser 
          Plasma Overwrap Thawing Bags (1000/Box) 
                    Plasma Bag Holder 
                    LABath CLEAN
TM
 
                    Reference Thermometer 
                    Thawing Bag Weights 
                    User's Manual 
 
 
Optional Accessories : 
                    Additional Bag Weights 
 
 
The above items may be purchased from : 
 
   GenesisBPS, LLC 
   866-712-5663 
 
 
2.    TO THAW BLOOD PRODUCTS 
 
2.1   
Place the unit on a flat, sturdy surface. 
 
2.2   Fill bath with water up to the slotted line on the front of the 
stainless steel shroud. This level will ensure proper thawing 
operation. No bags should be in the bath and the unit should 
be placed in the down position when filling the bath. 
 
 
Check water level daily. If the water level is too low, the 
GENESIS 2002/2006
 low water level sensing switch will 
make the unit inoperable (when the unit is plugged into the 
power supply). An alarm will sound and the ADD WATER