Test-Um VALIDATOR-NT NT955 User Manual

Page of 97
63
Test-Um NT955 Series
4.6. Step by Step Instructions for Performing Auto Tests
1.  From the Start Menu, navigate to the Job Manager using the 
arrow keys which move from left to right on each row.  
2.  Select the Job you wish to test.  Press Open.
3.  The Auto Test screen is presented with the Cable Test Schedule 
button highlighted. Press Select to view the Test Schedule.  
Press the up and down arrows to highlight the cable to be tested.
4.  Using the appropriate cable assembly and jacks, connect the 
Validator-NT main unit to one end of the cable and the remote 
unit to the opposite end of the first cable identified in the Test 
Schedule.
Note:  The Wiremapping Remotes (TP610) test wiremapping only and 
are NOT designed to work in the Auto Test mode.  The Test Results will 
indicate Fail because the Validator was not able to perform a complete 
test sequence.
5.  Press Test.  
6.  A message displays if the Main Unit is not communicating with 
the Remote Unit.  If the Remote unit is not found, the main unit 
automatically sends tone on the cable.  Using a Tone Tracer, 
such as Test-Um’s TT100, locate the cable you wish to test.  
Connect the Remote unit and testing will proceed.
7.  Pass/Fail Test Results display, also indicating the speed the 
cable is certified to perform at, and whether the cable will support 
Voice over IP (VoIP).  
8.  The Validator-NT automatically increments to the next cable to 
be tested. 
9.  Testing will proceed through the entire Cable Test Schedule 
regardless of test failures. 
Validator
TM
 Running Manual Tests
Validator
TM
 Running Manual Tests