M703E Calibrator Operator’s Manual 
M703E Calibration and Verification 
8.4.1. CALIBRATING THE PHOTOMETER’S SAMPLE GAS FLOW 
NOTE 
The procedure described in this section requires an independent, calibrated gas flow meter/monitor be 
connected to the EXHAUST fitting on the back of the M703E.   
During the 
PHOTO FLOW calibration, the M703E software automatically turns the DC pump downstream from 
the photometer 
ON.   
To perform a 
PHOTO FLOW calibration, press: 
 