30
Parameter Specification
Temp Coefficient
< 0.05% per deg C
Voltage Coefficient
< 0.05% per Volt AC (RMS) over range of nominal
10%
Dimensions (H x W x D)
7” x 17” x 23.5” (178 x 432 x 597 mm)
Weight
28 lbs (12.7 kg)
30.6lbs. (13.8kg) with IZS Option
30.6lbs. (13.8kg) with IZS Option
Environmental Conditions
Installation Category (Over voltage Category) II Pollution Degree 2
Table 2-2: IZS Generator Specifications with Reference Feedback
Option
Parameter Specification
Maximum Concentration
1.0 PPM
Minimum Concentration
0.050 PPM
Resolution 0.5
ppb
Repeatability (7 days)
1% of reading
Initial Accuracy
+/- 5% of target concentration
Response Time
< 5 min to 95%
Table 2-3: IZS Generator Specifications w/o Reference Feedback
Option
Parameter Specification
Maximum Concentration
1.0 PPM
Minimum Concentration
0.050 PPM
Resolution 0.5
ppb
Repeatability (7 days)
2% of reading
Initial Accuracy
+/- 10% of target concentration
Response Time
< 5 min to 95%
06870C DCN6332