Teledyne Microscope & Magnifier GFC7000E User Manual

Page of 265
Model GFC7000E Instruction Manual 
Operating Instructions 
04584 Rev A1 
111 
 
To enable automatic COM port reporting, follow the instruction shown in section 6.12.2.2 then 
press
 
 
 
 
 
Edit Data Channel Menu
 
 
 
SETUP X.X     
  0) CONC:     ATIMER,     1,          900
 
 
 PREV  NEXT            INS    DEL
   EDIT    
 PRNT 
   EXIT 
Exits to the main 
Data Acquisition 
menu 
 
 
 
SETUP X.X      
NAME:CONC
 
 
    <SET  
  SET> 
 EDIT
 
   PRINT 
 
EXIT
 
 
 
 
SETUP X.X      
RS-232 REPORT: OFF
 
 
    <SET    SET>
  EDIT 
   PRINT 
 
EXIT
 
ENTR accepts the new 
setting and returns to the 
previous menu. 
EXIT ignores the new setting 
and returns to the previous 
menu. 
SETUP X.X        
RS-232 REPORT: OFF
 
  
 OFF   
    
 
 
 
    ENTR    EXIT
 
 
Toggle key to turn 
reporting ON or OFF 
Press SET>  key until… 
From the DATA ACQUISITION menu 
(see Section 6.12.2.2)
 
 
6.12.2.8. Compact Report 
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of 
reporting each parameter in one channel on a separate line, up to five parameters are reported in 
one line. 
6.12.2.9. Starting Date 
This option allows to specify a starting date for any given channel in case the user wants to start 
data acquisition only after a certain time and date. If the Starting Date is in the past, the iDAS 
ignores this setting. 
6.12.2.10. Disabling/Enabling Data Channels 
Data channels can be temporarily disabled, which can reduce the read/write wear on the disk-on-
chip. The ALL_01 channel of the MGFC7000E, for example, is disabled by default.