Teledyne Microscope & Magnifier GFC7000E User Manual

Page of 265
Model GFC7000E Instruction Manual 
TROUBLESHOOTING & REPAIR PROCEDURES 
04584 Rev A1 
185 
 
Table 11-1:   Warning Messages - Indicated Failures 
WARNING 
MESSAGE 
FAULT CONDITION 
POSSIBLE CAUSES 
BENCH TEMP 
WARNING 
The optical bench temp 
is controlled at 48 ± 5
 
°
C. 
Bad bench heater  
Bad bench temperature sensor 
Bad relay controlling the bench heater 
Entire relay board is malfunctioning 
I
2
C
 buss malfunction 
BOX TEMP 
WARNING 
Box Temp is < 5 
°
C or 
> 48 
°
C. 
NOTE: Box temperature typically runs ~7
o
c warmer than 
ambient temperature.   
Poor/blocked ventilation to the analyzer. 
Stopped exhaust-fan 
Ambient temperature outside of specified range  
CANNOT DYN 
SPAN 
Dynamic Span 
operation failed 
Measured concentration value is too high or low. 
Concentration slope value to high or too low 
CANNOT DYN 
ZERO 
Dynamic Zero 
operation failed 
Measured concentration value is too high. 
Concentration offset value to high. 
CONFIG 
INITIALIZED 
Configuration and 
Calibration data reset 
to original Factory 
state. 
Failed disk on chip 
User erased data 
CONC ALRM1 
WARNING
 
Concentration alarm 1 
is enabled and the 
measured CO
2
 level is 
≥ the set point. 
 
CONC ALRM2 
WARNING
 
Concentration alarm 2 
is enabled and the 
measured CO
2
 level is 
≥ the set point. 
 
DATA 
INITIALIZED 
Data Storage in iDAS 
was erased 
Failed disk on chip 
User cleared data 
FRONT PANEL 
WARN 
The CPU is unable to 
Communicate with the 
Front Panel Display 
/Keyboard 
Warning only appears on serial I/O com port(s) 
Front panel display will be frozen, blank or will not respond. 
Failed keyboard 
I
2
C
 buss failure 
Loose connector/wiring 
PHOTO TEMP 
WARNING 
PHT DRIVE is  >2500 
mVDC 
Failed IR photo-detector 
Failed sync/demod board  
IR photo-detector improperly attached to the sample 
chamber  
Bench temp too high. 
REAR BOARD 
NOT DET 
Mother Board not 
detected on power up. 
 
Warning only appears on serial i/o com port(s) 
Front panel display will be frozen, blank or will not respond. 
Massive failure of mother board 
RELAY BOARD 
WARN 
The CPU cannot 
communicate with the 
Relay Board. 
I
2
C
 buss failure 
Failed relay board 
Loose connectors/wiring 
SAMPLE FLOW 
WARN 
Sample flow rate is < 
500 cc/min or > 1000 
cc/min. 
Failed sample pump 
Blocked sample inlet/gas line 
Dirty particulate filter 
Leak downstream of critical flow orifice 
Failed flow sensor/circuitry