Input and Output Calibration for Recorder with Display 
12/03 
DR4300 Circular Chart Recorder Product Manual 
141 
5.  Input and Output Calibration for Recorder with Display 
5.1 Overview 
Introduction 
This section explains how to field calibrate analog inputs, as well as a 4 to 20 mA current output for a model 
having a display and keypad.  
Because the recorder stores both factory and field calibration values for inputs in memory, you can easily 
select either factory or field calibration values for use by the recorder using the procedure given in 
Subsection 4.5.2.5.       
To calibrate an input: 
•  Determine the minimum and maximum range values to be used for calibration (see Subsection 5.2). 
•  Prepare to calibrate (see Subsection 5.3). 
•  Set up and wire the input for calibration (see Subsection 5.4). 
•  Use the parameters in the calibration set up group to store field calibration values in the recorder’s 
memory (see Subsection 5.5). 
What’s in this section? 
The following is a list of the topics covered in this section. 
Topic See 
Page 
5.1  Overview 
5.2  Input Calibration Minimum and Maximum Range Values 
5.3  Input Calibration Preliminary Information 
5.4  Input Calibration Set Up and Wiring 
5.5  Input Calibration Procedure 
5.6  Current Output Calibration 
 
  WARNING—SHOCK HAZARD   !
 
INPUT CALIBRATION MAY REQUIRE ACCESS TO HAZARDOUS LIVE 
CIRCUITS, AND SHOULD ONLY BE PERFORMED BY QUALIFIED 
SERVICE PERSONNEL.  MORE THAN ONE SWITCH MAY BE REQUIRED 
TO DE-ENERGIZE UNIT BEFORE CALIBRATION.