Hitachi Microscope & Magnifier S-4800 User Manual

Page of 294
3 - 8
3.2.3    Adjustment of Specimen Height 
 
Put the specimen stub on the specimen holder and adjust it to the proper height using the 
specimen height gauge as shown in the figure. To adjust, loosen the lock screw and adjust the 
specimen height so that the highest point of the specimen is the same as the bottom of the height 
gauge. Then, tighten lock screw. 
 
 
 
 
 
 
 
3.2.4  Specimen Exchange Position 
 
(1)   Type 1 (Manual stage): 
Confirm that the Stage Lock is free (the lamp of Stage Lock switch is off). 
Then set the XYRT and Z knobs of the specimen stage to the specimen exchange position. 
Specimen exchange positions: X: 25.0 mm 
Y: 25.0 mm 
     
 
 
R: 0゜   
T: 0゜   
Z: 8.0mm 
 
 
CAUTION
 
 
Specimen height must be adjusted carefully. It must not be 0.5mm higher than the bottom of 
the height gauge. If it is higher than this, the specimen may strike the objective lens and 
cause damage when operating at a short working distance or at a high tilt angle. Also, 
accurate setting of specimen height minimizes image shift during specimen tilting.
 
stub
specimen 
lock screw 
height gauge 
Y
R
Z
stage lock