Hitachi Microscope & Magnifier S-4800 User Manual

Page of 294
3 - 19
brightness in the field of view.    In such a case, decrease the size of the specimen, 
use a longer WD, or lower the accelerating voltage in order to correct the 
astigmatism. 
 
(7) Degauss Operation 
The Degauss operation eliminates hysteresis of the magnetic field in the objective lens.   
When focus is changed greatly, accuracy of magnification or alignment of the electron optical 
axis may degrade due to hysteresis of the focusing magnetic field.    Click Degauss button in 
the Column Tab under the following condition:   
    After changing focus widely.   
    Before making the electron optical axis alignment.   
Degaussing is automatically effected when WD is changed in the Column Tab or OPE-
CONDITION Block, when the accelerating voltage is changed, or when the Probe current 
mode is changed.   
 
(8) Setting of ABCC Link 
Used for setting whether or not to automatically adjust brightness and contrast (ABCC) upon 
changing the following status.     
    (a) Probe Current mode 
    (b) Specimen bias voltage 
  (c) Detector setting   
    (d) Magnification mode (High Mag/Low Mag) 
 
 
3.4.2.2  Low Mag Mode 
 
Probe Current mode and Cond Lens1 are set as same
 manner 
as in the HIGH MAG MODE area.   
These are selectable independently to High mag mode because in general the Low mag mode 
brings less bright image if operated with the same column condition.