Hitachi Microscope & Magnifier S-4800 User Manual

Page of 294
 
4-
17
4.4.10    When S-4800 control program does not start up 
 
When the progress bar on the login dialog window does not progress to the end, or the login 
dialog window disappear before completing start up, check the follows. 
 
(1) Check if the evacuation system is running.   
 
(2) Check if the Stage power switch is ON (for Type2 only). 
 
(3) Check that the HV cable is put correctly to the top of the electron gun. A safety protection 
switch is put at the part. It is to prevent electric shock when a user removed it while HV is ON. 
The switch is set ON automatically when the HV cable and safety cover are put correctly . If 
the switch is not ON, power of HV power supply is not provided and cause troubles for starting 
up. 
 
 
4.4.11    When PC has Hanged up 
 
When PC has hanged up while operating microscope, close windows and shut down the 
computer. 
It is recommended to shut the POWER switch of Display unit before starting up the S-4800 
program again.   
 
(1) Press Alt, Ctrl and Delete keys simultaneously. 
 
(2) Close Program dialog window appears.    Select a program which is shown as [Not 
responding] on the dialog window and click the End Task button.    If no program is shown as 
[Not responding], select a most doubtful one for hanging up.    If a message [This program is 
not responding, It may be busy, ......] appears, click the End Task button on it. 
If the computer recovered to normal state, you can start operation again.    If the computer did 
not recovered to normal state, press Alt, Ctrl and Delete keys again.    Click the Shut Down 
button on the Close Program dialog window.    The computer will be shut down. 
After a message [ Its now safe to shut your computer ] appeared, turn the power switch of the 
computer off. 
If shutting down by Alt, Ctrl and Delete keys is not effective, turn the power switch of the 
computer off confirming that the hard disk access lamp is not lit. 
 
(3) Start the computer again.    If a message [Windows was not properly shut down, .......Press 
any key to run Scan Disk on these drives......] appeared, press C key to run Scan Disk.    If 
Windows is started up in safe mode, shut down the computer using Start - Close Windows 
command on the taskbar.    Then start the computer again.