6.3    Specification of S-3400N External Communication (Option) 
6 - 18 
 
 
Acceleration voltage setting 
 
 
SetVacc ( Vacc as Long) as Long 
 
 
 
Vacc 
300 to 30000 :    0.3 – 30.0 (kV)   
  0.1kV unit at 10 to 30kV 
  0.01kV unit at 0.3 to 9.99kV 
 
 
 
Return 
0: Normal receiving 
 
 
 
 
Excluding 0: Abnormal receiving 
 
 
 
Comment: 
The method is finished at completion of 
command analysis. 
The completion of acceleration voltage 
setting is notified by the VaccFinished event. 
If the return value is not 0, the VaccFinished 
event is not issued. 
Please do not use other method until 
receiving the VaccFinished event. 
 
 Event 
  HV 
changing 
 
 
HVOnChange (HVOn as Long) 
   
HVOn  0: 
HV–OFF 
   
 
1: 
HV-ON 
 
 
 
Acceleration voltage changing 
 
 
VaccChange (Vacc as Long) 
   
Vacc 
0: 
HV–OFF 
 
 
 
 
300 to 30000 :    0.3 – 30.0 (kV) 
 
 
 
Comment: 
The event is not generated during the 
voltage increasing or decreasing.