Henkel LOCKTITE 97211 User Manual

Page of 60
12 
 
2    Description 
2.2    Displays, Operating Elements and Connections 
 
 
 
 
LED Row 1-10 V OUTPUT 
Displays the measured and amplified signal from the sensor during a measuring. The 
best adjustment of the amplification is during a measuring (dispensing) the signal will be 
not higher than 5 V. It has to be adjusted at the potentiometer 3
At the beginning of a measuring cycle the signal increases more than the adjusted 5 V. 
It shows you that the pressure sensor works well. If the sensor is damaged, no signal will 
be displayed. 
 
4 pin Plug
 
for connection to controller
 
The supplied connection cord is connected here (Pin assignments see Section 8.2). 
 
Potentiometer R10
 
to adjust the amplification factor
 
It is necessary to adjust the amplification factor for getting an optimum signal for the 
interpretation at the Loctite controller. 
 
Service Socket,
 
for Loctite service only
 
 
4 pin Socket
 
for connection the pressure sensor
 
The cable of the pressure sensor is connected here (Pin assignments see Section 8.2). 
1
2
3
4
5
Loctite (Ireland) Ltd.
Made in Germany
cat. no. 97211
Tallaght Business Park
Whitestown
Tallaght, Dublin 24, Ireland