Extech Instruments Film Camera 380460 User Manual

Page of 5
380460/380462-EU-EN v2.4   10/13 
Specifications 
General Specifications 
Circuit 
Custom one-chip LSI microprocessor circuit 
Display  
18 mm (0.7") LCD with 2000 display counts 
Measurement terminals 
4-Terminal Kelvin type 
Measurement ranges 
Five (5) ranges (200m, 2, 20, 200, 2000
 ohms) 
Zero Adjust 
±50 count adjustment 
Sampling Time 
Approximately 0.4 seconds 
Over input indication 
Indication of "1  " 
Operating  Temperature  
0 °C to 50 °C (32 °F to 122 °F) 
Operating Humidity 
80% Relative Humidity max. 
Power Supply 
110V (380460) or 220V (380462) ±15%, 50/60Hz  
Power Consumption 
Less than 2 VA 
Weight       
694g (1.53 lbs.) 
Dimensions    
167 x 122 x 85 mm (6.6 x 4.8 x 3.4") with cover 
Accessories 
Power Cable and 4-wire Kelvin clip leads 
Range Specifications 
 
Range 
 
Resolution
Test 
Current 
 
Accuracy 
Open Circuit
Voltage 
200 m
 
0.1m
 
100mA 
± 0.75% + 4 digits
4.4V 
 
1m
 
10mA 
± 0.75% + 2 digits
3.6V 
20 
 
10m
 
10mA 
± 0.75% + 2 digits
3.6V 
200 
 
0.1 
 
1mA 
± 0.75% + 2 digits
3.5V 
2000 
 
 
1mA 
± 0.75% + 2 digits
3.5V 
Note: Specifications tested using RF Field Strength <3V/m and frequency <30MHz 
 
 
 
 
 
 
Copyright © 2013 FLIR Systems, Inc. 
All rights reserved including the right of reproduction in whole or in part in any form. 