Emerson 3812 User Manual

Page of 138
Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
Section 3: Electrical installation
3-9000-765 Rev D
March 2013
Frequency/Digital outputs
 59
Output for FODO1 and Digital Output1 (Group 1 on the CPU Module) share a common ground 
and have 50V isolation. FODO2 and FODO3 (Group 2 on the CPU Module) share a common 
ground and  have 50V isolation. This allows an output  to be connected to a different flow 
computer. The outputs are opto-isolated from the CPU Module and have a withstand voltage of 
at least 500V rms dielectric.   
Figure 3-6  CPU Module - Frequency/Digital outputs common ground 
A.
A. FODO1 and Digital input1 - shared common ground (Group 1)
B. FODO2 and FODO3 - shared common ground (Group 2)
B.