Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
Section 3: Electrical installation
3-9000-765 Rev D
March 2013
Direct or remote mount transmitter electronics enclosure seal
 63
Procedure
1.
Rotate the end cap clockwise fully closing and compressing the end cap seal. Install the 
Security latch using a 3mm Allen wrench.
2.
Install the security seal wire into and through one of the two holes in the end cap. 
Choose holes that minimize counterclockwise rotation of the end cap when the 
security wire is taut (maximum wire diameter .078 inch; 2.0mm). 
Figure 3-9  Direct or remote mount Transmitter Electronics Enclosure security seals 
3.
Adjust the security wire, removing all slack and thread into the lead seal. 
4.
Cut wire ends to remove excess wire.
5.
This completes the direct or remote Transmitter Electronics Enclosure seal installation 
procedure.
A.
A.  Transmitter Electronics Enclosure end cap
B.  Security wire seals 
B.