Chapter 1
265
GPIB Command Dictionary
MEASure subsystem
:DEC:DTMF:SYMBol?
These command set/query the symbol number assigned by the encoder
for each of the DTMF tones. As each tone is analyzed, the symbol that
represents each tone is returned as a quoted string.
for each of the DTMF tones. As each tone is analyzed, the symbol that
represents each tone is returned as a quoted string.
Syntax
MEAS:DEC:DTMF:SYMB? !returns a quoted string
Screen/field equivalent
MEAS:DEC:DTMF:SYMB? reads the
Sym
column on the SIGNALING
DECODER screen when the
Mode
field is set to
DTMF
.
:DEC:FGENerator:FREQuency <meas cmnd>
:DEC:FGENerator:FREQuency?
:DEC:FGENerator:FREQuency?
These commands set/query the frequency of the signal generated by the
function generator.
function generator.
Syntax
MEAS:DEC:FGEN:FREQ <meas cmnd>
MEAS:DEC:FGEN:FREQ? !returns and integer value
Screen/field equivalent
MEAS:DEC:FGEN:FREQ? reads the
Frequency
measurement field on
the SIGNALING DECODER screen when the
Mode
field is set to
Func
Gen
.
:DEC:NAMPs|NTACs:NBITs <meas cmnd>
:DEC:NAMPs|NTACs:NBITs?
:DEC:NAMPs|NTACs:NBITs?
These commands set/query the number of bits measured by the
signaling decoder in NAMP-NTAC mode.
signaling decoder in NAMP-NTAC mode.
Syntax
MEAS:DEC:NAMP:NBIT <meas cmnd>
MEAS:DEC:NAMP:NBIT? !returns and integer value
Screen/field equivalent
MEAS:DEC:NAMP:NBIT? reads the
Num of Bits
measurement field
on the SIGNALING DECODER screen when the
Mode
field is set to
NAMP-NTAC
.