Chapter 1
265
GPIB Command Dictionary
MEASure subsystem
:DEC:DTMF:SYMBol?
These command set/query the symbol number assigned by the encoder 
for each of the DTMF tones. As each tone is analyzed, the symbol that 
represents each tone is returned as a quoted string.
Syntax
MEAS:DEC:DTMF:SYMB? !returns a quoted string
Screen/field equivalent
MEAS:DEC:DTMF:SYMB? reads the 
Sym
 column on the SIGNALING 
DECODER screen when the 
Mode
 field is set to 
DTMF
.
:DEC:FGENerator:FREQuency <meas cmnd>
:DEC:FGENerator:FREQuency?
These commands set/query the frequency of the signal generated by the 
function generator.
Syntax
MEAS:DEC:FGEN:FREQ <meas cmnd>
MEAS:DEC:FGEN:FREQ? !returns and integer value
Screen/field equivalent
MEAS:DEC:FGEN:FREQ? reads the 
Frequency
 measurement field on 
the SIGNALING DECODER screen when the 
Mode
 field is set to 
Func 
Gen
.
:DEC:NAMPs|NTACs:NBITs <meas cmnd>
:DEC:NAMPs|NTACs:NBITs?
These commands set/query the number of bits measured by the 
signaling decoder in NAMP-NTAC mode.
Syntax
MEAS:DEC:NAMP:NBIT <meas cmnd>
MEAS:DEC:NAMP:NBIT? !returns and integer value
Screen/field equivalent
MEAS:DEC:NAMP:NBIT? reads the 
Num of Bits
 measurement field 
on the SIGNALING DECODER screen when the 
Mode
 field is set to 
NAMP-NTAC
.