8.   CROSSLINKER CALIBRATION PROCEDURE 
 
Equipment necessary: 
 
Chart recorder 
Two probes with banana plug adapters 
One radiometer with UVX-25 calibrated sensor 
Cable with banana plug and jack adaptor to connect radiometer to chart-recorder 
Power source (100V, 115V or 230V) 
 
Procedure: 
 
1.  Connect the crosslinker to the power source. Do not turn on the unit at this time. 
 
2.  Connect the testing probes to the chart recorder and to the crosslinkers PCB as follows: 
 
 
PEN 1 (red) to the right side of R7 
 
 
PEN 2 (blue) to the right side of R6 
 
 
Ground both pens at the lower part of R2 
 
Note: while connecting these probes, the switches for both pens on the chart recorder 
should be on ZERO position to avoid possible damage to the recorder. 
 
3.  Place the UVX-25 sensor inside the crosslinker at the center and slide it through the upper 
hole of the inner chassis. Plug the sensor into the PCB's jack. 
 
4.  Select 5V scale for both pens. Turn the chart recorder on. 
 
5.  Select a speed of 15mm/min. 
 
6.  Switch both pens to MEASURE and DOWN positions so they can touch the chart paper. 
 
7.  Adjust the ZERO position for both pens. 
 
8. Press 
the 
CHART DRIVE switch. 
 
 
 
 
 
 
 
 
 
 
 