Disabling the HR Alarm
4-7
4
Disabling the HR Alarm
If the heart rate is outside the HR alarm limits, an alarm sounds. To disable the
alarm, press
.
appears to indicate that the alarm is disabled.
Adjusting the ECG Size
To increase or decrease the size of the ECG, press
or
on the gain con-
trol, .
Troubleshooting
Table 4-5 provides troubleshooting tips for ECG Monitoring.
+5$/$50
Table 4-5 Troubleshooting Tips
Situation
Cause
Solution
/HDGV2II
message
or
dashed line (------)
z
The monitoring elec-
trodes are not applied
or are not making
proper contact with the
patient.
trodes are not applied
or are not making
proper contact with the
patient.
z
The monitoring cable is
not connected.
not connected.
z
Check that the monitoring
electrodes are properly
applied.
electrodes are properly
applied.
z
Check that the monitoring
cable is properly con-
nected.
cable is properly con-
nected.
3DGV2II
message
z
The pads are not mak-
ing proper contact with
the patient.
ing proper contact with
the patient.
z
Check that the pads are
properly applied.
properly applied.