Philips HEARTSTART XLT M3500B User Manual

Page of 180
Delivering a Synchronized Shock
7-3
7
Figure 7-1:  Charging in Sync Mode
If desired, you may increase or decrease the selected energy level after 
pressing 
by pressing 
 repeatedly until the desired energy 
level is displayed. The defibrillator charges to the modified energy level 
automatically. Wait until the current charge reaches the selected energy 
level before proceeding. 
5. Make sure no one is touching the patient or anything connected to the 
patient. Call out “Clear.” Then press 
 and continue to hold 
 
down until the shock is delivered. 
The defibrillator shocks with the next detected R-wave.
WARNING
Defibrillation current can cause operator or bystander injury. Do not touch the 
patient, or equipment connected to the patient, during defibrillation. 
ENERGY
SELECT
HR ALARM
LEAD
SELECT
DISARM
Shocks:0
00:00:8
78
Pads
SpO2
150 J
87 J
Sync
&+$5*(
(1(5*<
6(/(&7
6+2&.
6+2&.