Calibrating I/O
Calibrating Analog Channels
Analog Input can be calibrated for Analog and Temperature IOLAN models. 
Select the channel you want to calibrate. This example uses an A4 model that has channel A1 set to 
Current with a Range of 0 to 20mA.
If you have not disabled confirmation messages (
Tools
Options
 in DeviceManager only), you will 
get prompted to verify channel calibration.
Click 
Yes
 to proceed with calibration. You are now prompted to apply 0 mA to the positive (+) and 
negative (-) terminals. Once that is done, click 
Yes
 to proceed.
You are now prompted to apply 20 mA to the positive (+) and negative (-) terminals. Once that is 
done, click 
Yes
 to proceed.
Once calibration is successfully completed, click 
OK
 to finish the process.
Resetting Calibration Data
You can reset the I/O channels calibrations to the factory calibrations in DeviceManager by selecting 
Tools
I/O Channels
Reset Calibrate Data
 or in WebManager by selecting 
Administration
Reset
I/O Calibration
.