Connections and Wiring 
   
The DX-ARC-SWB connections are located along the front panel of the case. 
When all wiring has been completed, it must be securely fastened to the 
powerchair frame to ensure there is no strain on the connectors. 
 
Ensure that all wiring is suitably restrained to prevent snagging. 
Ensure that the unit and all connections are suitably protected against 
environmental contamination (water, dust, etc.). 
Before making any connections to the DX-ARC-SWB, disable the 
powerchair by one of the following means to prevent accidental 
movement: 
1) Place the battery circuit breaker in the open position, 
2) Disconnect the motor or batteries, and/or 
3) Elevate the drive wheels. 
 
4.1 DXBUS 
The DX-ARC-SWB has 2 DXBUS sockets to connect to the DXBUS system. Both 
sockets are identical. Connect a DXBUS cable from either socket to the DXBUS 
socket of any other DX Module. Route and secure the cable to prevent mechanical 
damage and snagging.  
The second socket can be used to extend the DXBUS to other DX Modules, if 
required. If it is not used for this purpose, and the vacant socket is accessible to the 
user, a dummy plug or cap should be fitted.