8.0 CALIBRATION 
 
•  Calibration may be required when the balance is initially installed or 
if the balance is moved from a location.      
 
•  With time and use, mechanical deviations may occur.  Therefore 
calibrate the balance whenever the best accuracy is required or 
when any deviation is observed during weighing a test weight.    
 
•  Before calibrating it is necessary to have the proper calibration 
weights.  The weights should be known to an accuracy that is 
appropriate for the balance being calibrated, for example, OIML 
Class M1 type or ASTM E617 Class 4.  If you do not have proper 
weights do not attempt calibration.   
 
Model  
AQT-200 
AQT- 250 
AQT-600 
AQT-1500 
AQT-2600 
AQT-2000 
AQT-5000 
Calibration 
Weight 
100g 
Or 
200g 
200g 
Or 
250g 
500g  
Or 
600g 
1000g  
Or 
1500g 
1000g  
Or 
2000g 
1000g 
Or 
2000g 
2000g 
Or 
5000g 
 
8.1  AUTO CALIBRATION (FOR AQT 200 ONLY) 
 
•  Before entering calibration mode, press [On/Off] to turn the power 
on for 1 minute. 
 
Enter calibration mode 
 
• Press [On/Off] to turn the power off.  
• Press [Unit] first and then press [On/Off], while keeping the [Unit] 
key pressed. The display will show the A/D value.  Then press [Unit] 
and the display will show the last calibration value used, press [Tare] 
to choose either 100g or 200g.  
 
Calibration  
•  Wait for the stable indicator to be displayed,  
• Press [Unit]  - the display will show “L 100" or “L 200”, with the "L" 
flashing.  
Place the appropriate weight on the platform.  A beep will be heard 
after a few seconds and the display will show the weights value.    
• Press [On/Off], to turn off the power.  
• Press [On/Off] to turn on the power again. Place maximum weight 
(200g) on the platform to verify weighing is correct. If not, repeat the 
calibration. 
 
 