Agilent Technologies 8156A User Manual

Page of 289
 127
Remote Commands
User Calibration Commands
The value that you send with this command, is 
the attenuation for the next calibration point. 
The wavelength of the calibration point is 
updated automatically. The first piece of data is 
for the start wavelength specified by the 
:UCAL:START
 command. The default value 
for the value is dB.
The value can be in the range 0.001dB to 
99.999dB.
:UCALibration:VALue? 
Syntax    
:UCALibration:VALue?
Description    
The query returns a value from the user 
wavelength calibration data.
The value returned is the attenuation for the 
next calibration point. The wavelength of the 
calibration point is updated automatically. The 
first piece of data is for the start wavelength as 
returned by the 
:UCAL:START?
 query. The 
values returned are in dB.
The error 204 indicates that there are no more 
data points to be read.