136
Startup page
Section 5: Startup and meter configuration
Reference, Installation, and Operations Manual
June 2013
3-9000-743 Rev S
To enable / identify the source of temperature / pressure for meter corrections,
configure these data points listed in the table below (requires the Option Board):
configure these data points listed in the table below (requires the Option Board):
5.
Click the checkbox(s) to enable or disable the desired Meter Outputs:
•
Frequency outputs
•
Current outputs
•
HART® outputs (if the Series 100 Plus Option Board is installed)
•
Digital outputs
6.
Click the checkbox(s) to enable or disable:
•
View Gas Chromatograph Setup
•
View Continuous Flow Analysis
7.
Click the
Next
button to continue the meter configuration. Daniel MeterLink displays
the
Field Setup - General Page.
Pressure expansion
correction
correction
Data points affected:
•
EnableExpCorrPress
Options:
•
Disable (FALSE)
•
Enable (TRUE)
Guidelines:
•
For options and guidelines, see Daniel MeterLink online help.
Table 5-13 Enable Temperature/Pressure Meter Corrections Data Points, Source
Daniel MeterLink
Display Name
Display Name
Data Points, Options and Guidelines
Temperature for meter corrections
Data points affected:
- EnableTemperatureInput
Options:
•
Not used (0)
•
Live (1) (only available with the Option Board)
•
Fixed (2)
Guidelines:
•
For options and guidelines, see Daniel MeterLink online help.
Pressure for meter
corrections
corrections
Data points affected:
- EnablePressureInput
Options:
•
Not used (0)
•
Live (1) (only available with the Option Board)
•
Fixed (2)
Guidelines:
•
For options and guidelines, see Daniel MeterLink online help.
Table 5-12 Enabling Meter Corrections
Daniel MeterLink
Display Name
Data Points, Options and Guidelines